How to Measure Electrical Conductivity (EC) Correctly?

The topic of this article is proven methods for measuring the electrical conductivity of a solution, a value that plays a very important role especially in indoor plant cultivation. If you are interested in how to use EC metres, or EC/TDS probes, or how to care for them properly so that your measuring instruments last as long as possible, you are in the right place.

a_EC_COVER1

If you would first like to learn what EC actually is, you can start with a separate article: Electrical conductivity of a solution – EC. In practice, this means determining the amount of mineral (“nutrient”) salts dissolved in the nutrient solution.

Measuring EC frequently and accurately is an indispensable part of plant cultivation not only in a growbox. Many beginner growers then wonder what the difference is when measuring between EC (the measured value stated in μS/cm) and TDS (the measured value stated in ppm). The difference lies only in the fact that from the TDS value (Total Dissolved Solids) you can read the total amount of dissolved solids in ppm units.

EC metres and their calibration

EC calibration solutions are available in 1.413 mS and 2.76 mS variants. However, do not be misled by the 2.8 mS or 2.77 mS variants either, which are equivalent to 2.76 mS solutions and the only difference is rounding.

a_EC_CONDUCTIVITYKOMPLET

How often should you calibrate an EC metre?

The answer here is straightforward. To ensure maximum measurement accuracy, the electrodes should ideally be calibrated using the above calibration solutions before each use. If the probe is used on a daily basis, it should also be calibrated every day.

And now we come to further proven methods for measuring conductivity:

  • Always use a fresh calibration solution! In other words, calibration standards are easily contaminated and, moreover, they do not have buffering capacity, so there is no point in trying to use them repeatedly.
  • Before you start calibrating, rinse the probe with deionised water (see separate article: Reverse osmosis). Only then can you safely immerse the probe in the calibration standard. For even more accurate measurement results, prepare a second separate beaker with calibration solution or sample for rinsing in advance. Proper rinsing of the electrode with deionised water is truly very important. Accumulated salts and other residues can contaminate solution samples and cause inaccurate readings. Therefore, rinse the probe every time you remove it, before storage, or when measuring conductivity of different samples in multiple beakers.
  • Plastic beakers are generally recommended because other materials, such as glass or metal, cause electromagnetic interference with the measured electrical conductivity.
  • Correct placement of the electrode in the beaker means the same distance from the beaker wall in all directions.
  • Also make sure that the electrode is sufficiently immersed. Otherwise, you will only get distorted results when measuring conductivity. Therefore, check that all openings are actually below the surface and that there are no bubbles between the sensors.

a_EC_sondapříklad

Example of a replacement electrode for an EC metre.

  • Check for air bubbles, because air may remain trapped in the body, or on the body, of the electrode even when it is immersed, causing inaccurate measurement. If you notice air bubbles, immerse the probe sufficiently and move it up and down. You can then release any bubbles from the openings by gently tapping it.
  • Although electrodes usually have so-called temperature compensation, bear in mind that it takes them some time to reach thermal equilibrium, so always wait until the measured conductivity or temperature value stabilises.
  • Choose the appropriate type of electrode, because not all electrodes are suitable for the given range or sample. Sometimes, when selecting a measuring device, the suitable range as well as chemical compatibility may therefore matter. For certainty, you can always consult us (see below).
  • In addition to frequent calibration, regular cleaning is also a prerequisite for ensuring the longest possible service life and proper functioning of every EC electrode, although this is often underestimated. This is despite the fact that deposits can accumulate in the electrode section as well as on the vent openings of the probe, which is a common cause of calibration problems and inaccurate measurement results. So how should you clean the electrode properly? A proven method is to use warm soapy water, with which you thoroughly rinse the electrode. Do not use chemical cleaning solutions or solvents.

Some multifunction measuring instruments use so-called combined electrodes, and we discuss their proper storage at the end of a separate article entitled: What is the storage solution (KCL) for? For proper storage of the EC electrode, it is essential that in the case of a combined EC/pH electrode, improper storage may cause the electrolyte to dry out undesirably. Sometimes dried-out probes can be regenerated by immersion in storage solution for pH

 

 

If you are interested in working with pH metres, we have also prepared a separate article for you: How to use calibration solutions and pH metres?

If you would like to ask us anything, not only about the correct measurement of conductivity, we will be happy to advise you, for example at the well-known e-mail address info@higarden.eu.

We also recommend these products:

What to read next: